Introduction
Miniaturized structure, medical temperature sensing catheter inserted into the rectum or implanted subcutaneously to measure and record core body temperature, rechargeable and reusable, used in experimental animal temperature measurement and life medicine research.
Scope of application: This product measures the temperature of the rectum, oral cavity, and subcutaneous area. It is suitable for measuring the core body temperature of humans or animals, and is also particularly suitable for measuring the body temperature of infants and young children.
Technical parameter
Product model: SV227
Communication protocol: BLE4.2 or above
Recording interval: 5 minutes
Use time after one charge ≥ 72 hours
Working conditions: 0~50℃; ≤85%RH
Measuring range: 25.0℃~45.0℃
Resolution: 0.1℃
Measurement error: ± 0.2℃
Net weight: 1.3g
Waterproof: This product is sealed as a whole and has waterproof function.
Storage conditions: -25~55℃; ≤85%RH; no corrosive gas and well-ventilated room.
Validity period: 3 years
Operation steps
—Install the App software
The TempView application is only applicable to smartphones with Android operating systems, not to phones with Apple, Hongmeng, and other operating systems. Please download the program from the following address:
2 Scan the QR code with your phone and download our TempView App application in Google Play.
— Charging the thermometer
Put the thermometer back into the storage box immediately after use. Do not leave it outside the storage box, otherwise it may cause permanent damage to the thermometer! The storage box has a built-in battery to charge the thermometer, and it can be used for 72 hours each time the thermometer is taken out.
—Measure oral temperature
Put the thermometer probe into the root of the tongue under the tongue in the mouth, close your lips, and read the temperature measurement data after five minutes.
Step 1 → Step 2 → Step 3 → Step 4
Tear off one side of the double-sided adhesive sticker and stick it on the side of the thermometer. Insert the probe into the anus to a depth of ≥3cm. Stick the thermometer on the skin.
—Implant subcutaneously to measure body temperature
Implant the thermometer’s temperature probe subcutaneously to measure the animal’s body temperature, and stick the capsule to the animal’s body.
—Use transparent waterproof tape
According to the diagram below, use transparent waterproof tape to stick the thermometer to the skin to achieve the purpose of waterproofing, sweat-proofing, and not falling off.
Warning
· Do not modify this product in any form.
· Do not allow the thermometer to come into contact with chemical solvents or diluents.
· Do not squeeze the product hard. If there is any damage on the appearance, please stop using it immediately.
· This product cannot be used with MRI or CT equipment.
· This product is strictly prohibited to be used in an environment with flammable anesthetic gas.
Safety precautions
· This product only plays an auxiliary role in patient diagnosis. Please make a diagnosis based on clinical manifestations and symptoms.
· Please pay attention to the temperature and humidity requirements for the use of this product. Using it in an environment that does not meet these requirements will cause inaccurate temperature measurement results.
· The effective Bluetooth transmission distance of this product in an accessible environment is 10 meters.
· After using the thermometer, please put the thermometer back into the charger storage compartment in time to avoid affecting the service life of the product.
· Lithium batteries are dangerous. Users must not try to disassemble the battery. Forcible disassembly or replacement of the battery may cause an explosion. The battery may also emit a peculiar smell when thrown into a fire. Please do not use it. Please dispose of used waste batteries according to the manufacturer's instructions.
· If you feel uncomfortable (discomfort or allergic symptoms) when wearing this product, please stop using it immediately and consult a doctor.
· Please use this product with caution in ICU, strong magnetic environment, etc.
· Do not put the thermometer in liquid for cleaning or disinfection, and do not allow liquid to flow into the thermometer.
Maintenance and care
· When the APP reminds you that the thermometer battery is low, please charge it in time, otherwise it may not work properly.
· The handling of products and batteries at the end of their service life should comply with local laws and regulations.
· Cleaning: When the body and temperature sensing parts are dirty, use a clean soft cloth with a small amount of water or neutral detergent, wring it out and wipe it.
· Disinfection: Before and after use, wipe the thermometer with a clean soft cloth with a small amount of 75% concentration of medical alcohol and dry it naturally.
Packing list
Insertion thermometer SV227 1 unit
Storage box SC127 1 unit
Medical double-sided adhesive sticker 1 sheet (optional)
Medical transparent waterproof tape 10 sheets (optional)
SV227 user manual 1 copy
TempView App user download